Run SIP (Sterilization In Place).

POST 

/sip

Pre-filling pipeline SIP (Sterilization In Place):

1.Connect the Cleaning connector to the Nozzle.

2.Manually switch Valve-3 to the SIP position.

3.Confirm that Valve-2 is in the SIP/Dry position.

4.Confirm that Valve-1 is in the SIP position.

5.Control the Pump to perform the pre-filling pipeline SIP for Agar.

Post-filling pipeline SIP (Sterilization In Place):

1.Manually switch Valve-3 to the SIP position.

2.Manually switch Valve-2 to the SIP/Dry position.

3.Control the Pump to perform the post-filling pipeline SIP.